Abstract:Timdtemperature profiles for the Bausch and Lomb Low Heat Contact Lens Disinfection Unit were established during operation under a number of extreme but possible environmental conditions. These included low voltages, partially filled contact lens cases, low environmental temperature, grit in the heat well, saline in the heat well and combinations of these. Only the extreme combination of low temperature and grit in the heat well lowered performance to an unacceptable level. We recommend that practitioners stro… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.