The Influence of Flux on the Reliability of Solder-Sealed Hermetic Packages for Active Implantable Medical Devices

Abstract: This paper discusses the influence of fluxing materials on the properties of lead-free solder seals on screenprinted metal thick films. Pull tests revealed a strong impact on the layer adhesion if an aggressive flux is used. A milder flux was able to improve the pull test results slightly. Investigations on the hermeticity demonstrated the difficulty of hermetic packaging with SnAg solder. None of the fluxes under test allowed the manufacturing of hermetic packages in the desired range of 10 -9 mbar l s -1 or … Show more