The Influence of X-Ray Tube Current-Time Variations Toward Signal to Noise Ratio (SNR) in Digital Radiography: A Phantom Study

Abstract: Several image enhancement applications can be used to increase SNR and brightness. However, if the setting of tube current-time is too high, the application can not reduce the radiation dose. So, It is necessary to test tube current-time variations on the SNR value and image quality to determine the effect on body organs and radiation dose. This study aims to determine the influence of tube current-time on SNR and with this research is expected to be an input for radiographers in the field to pay attention to … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
3
0

Year Published

2023
2023
2023
2023

Publication Types

Select...
1
1

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(3 citation statements)
references
References 31 publications
0
3
0
Order By: Relevance
“…X-ray tube current (mA): The quantity of electrons generated by the cathode filament in the cathode assembly of the X-ray tube is determined by the tube current [44]. During X-ray generation, these electrons collide with the anode target of the X-ray tube to generate X-rays photons.…”
Section: How Exposure Factors Affect Image Qualitymentioning
confidence: 99%
See 1 more Smart Citation
“…X-ray tube current (mA): The quantity of electrons generated by the cathode filament in the cathode assembly of the X-ray tube is determined by the tube current [44]. During X-ray generation, these electrons collide with the anode target of the X-ray tube to generate X-rays photons.…”
Section: How Exposure Factors Affect Image Qualitymentioning
confidence: 99%
“…2. X-ray tube current (mA): The quantity of electrons generated by the cathode filament in the cathode assembly of the X-ray tube is determined by the tube current [44].…”
Section: Phantom Aluminum Platesmentioning
confidence: 99%
“…2. X-ray Tube Curent (mA): The quantity of electrons generated by the cathode filament in the cathode assembly of the X-ray tube is determined by the tube current [43]. During X-ray generation, these electrons collide with the anode target of the Xray tube to generate X-rays photons.…”
Section: How Exposure Factors Affect Image Qualitymentioning
confidence: 99%