“…Theoretically, LVOT-AR ranges from 0 to 1 (Figure 3). [2][3][4][5][6][7] Furthermore, the reproducibility of this VD method was evaluated by 2 local trained observers at Yamaguchi University Hospital and by core laboratory staff at an independent core frames are recorded -software required; (2) the contrast enhancement-dynamic mode is inactivated; (3) during image acquisition, the patient and the table of the cath-lab are kept immobile; (4) the tip of the pigtail catheter is located just above the leaflet coaptation level to avoid the interference of the catheter with the valve leaflets (Table; Figure 1); and (5) after identification of the optimal implanter view, in which the nadirs of 3 coronary cusps are aligned, using a dedicated software (3 mensio Valve version 8.0, 3 mensio Medical Imaging BV, Bilthoven, Netherlands), the optimal angiographic view for VD analysis is identified, in order to avoid the overlap of background structures with the ROI (sub-aortic segment designed by the basal third of the LV on fluoroscopy) and to minimize LVOT foreshortening on MSCT. Figure 2 shows examples of volume rendering images reconstructed using the dedicated software (3 mensio Valve version 8.0; 3 mensio Medical Imaging BV, Bilthoven, Netherlands; Figure 2A,A',D,D'), 2-D virtual "fluoroscopic" view derived from MSCT ( Figure 2B,B',E,E') and aortograms ( Figure 2C,C',F,F').…”