The Medtronic Sprint Fidelis® lead history revisited—Extended follow‐up of passive leads

Abstract: Background Due to high failure rates, Medtronic withdrew the Sprint Fidelis lead (SFL) from the market. Passive fixation lead models exhibited better survival than active models, but most studies have limited follow‐up. Aim of this study was to give insights into passive lead survival with a follow‐up of 10 years. Methods In two large Swiss centers, patients with passive SFLs were identified and data from routine implantable cardioverter defibrillator (ICD) follow‐ups were collected. Patients were censored at … Show more