The Multi-Blade Boron-10-based neutron detector for high intensity neutron reflectometry at ESS

Abstract: The Multi-Blade is a Boron-10-based gaseous detector introduced to face the challenge arising in neutron reflectometry at pulsed neutron sources. Neutron reflectometers are the most challenging instruments in terms of instantaneous counting rate and spatial resolution. This detector has been designed to cope with the requirements set for the reflectometers at the upcoming European Spallation Source (ESS) in Sweden. Based on previous results obtained at the Institut Laue-Langevin (ILL) in France, an improved de… Show more

Help me understand this report
View preprint versions

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

12
133
0
1

Year Published

2018
2018
2023
2023

Publication Types

Select...
8

Relationship

1
7

Authors

Journals

citations
Cited by 45 publications
(148 citation statements)
references
References 63 publications
12
133
0
1
Order By: Relevance
“…In average each detector pixel records several counts per second, see figure 8, the operation regime for both the Multi-Blade and the He-3-based AMOR detector is below any saturation either due to electronics limitations or physical detector limitations. Indeed, the actual limits of the Helium-3 technology is around hundreds Hz/mm 2 [23], and the Multi-Blade detector is expected to achieve the order of kHz/mm 2 [2]. Therefore, in this configuration, the total rate is recorded by the detector and it can be used for the comparison with the local instantaneous peak rate measurement.…”
Section: Counting Rate Capabilitymentioning
confidence: 99%
See 2 more Smart Citations
“…In average each detector pixel records several counts per second, see figure 8, the operation regime for both the Multi-Blade and the He-3-based AMOR detector is below any saturation either due to electronics limitations or physical detector limitations. Indeed, the actual limits of the Helium-3 technology is around hundreds Hz/mm 2 [23], and the Multi-Blade detector is expected to achieve the order of kHz/mm 2 [2]. Therefore, in this configuration, the total rate is recorded by the detector and it can be used for the comparison with the local instantaneous peak rate measurement.…”
Section: Counting Rate Capabilitymentioning
confidence: 99%
“…The proof of concept for this detector technology has been primarily demonstrated with a first test of a prototype at the ILL in 2013 [1]. A detailed characterization [2] of a Multi-Blade detector has been carried out at the Budapest Neutron Centre [35][36][37] in 2015. In 2017 a reflectometry demonstrator has been installed and tested at the neutron reflectometer CRISP [38] at the ISIS neutron and muon source in the UK [39][40][41] performing a full detector characterization along with specular and offspecular reflectivity measurements on several samples [3,4].…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…Recently, new position-sensitive neutron detectors (PSNDs) with solid layers containing 10 B have been created. They are hybrid detectors [1][2][3], which contain 10 B layer used as a converter of neutron into charged particles which are detected in a gas-discharge chamber. The advantage of such detectors to detectors with helium-3 is an ability to operate in a wide range of neutron energies.…”
Section: Introductionmentioning
confidence: 99%
“…The planned ESS instrument suite challenges the state of the art of the position sensitive neutron detection by requesting detectors able to sustain count rates which may reach the MHz range [1] (a 3 He proportional counter can sustain up to 150 kHz [2]) at a finer spatial resolution (from the micrometric range to some mm) over relatively large areas (up to tens of m 2 ). Current scientific CMOS camera sensors, considerably more robust and cheaper than ECCD, have reached sensitivities of the order of a few photons.…”
Section: Introductionmentioning
confidence: 99%