The phenomenon of drop in output at larger field sizes for telecobalt units

Abstract: It is known that the output factors (OPFs) for external-beam radiotherapy units increase with field size due to increased scattered radiation from the collimator system. Saturation in the OPF value is generally reported beyond approximately 30 × 30 cm(2). For the first time, to the best of our knowledge, we report on a drop in OPF values, although marginal, measured for a telecobalt machine beyond the 38 × 38 cm(2) field size. We believe that reporting and explaining the results will lead to a better understan… Show more