Abstract:BackgroundPower‐on reset (PoR) is most commonly due to electromagnetic interference. Full PoR results in a switch to an inhibited mode (VVI) pacing and resets pacing outputs to maximal unipolar settings, leading to extracardiac stimulation.MethodsWe present a case of PoR occurrence in the absence of electromagnetic interference, resulting in pectoral stimulation triggered by violation of the atrial rate limit.ConclusionsIt is useful for clinicians to recognizethe occurrence of PoR in the setting of atrial limi… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.