The routine measurement of platelet volume: a comparison of light-scattering and aperture-impedance technologies

Abstract: The effect of dipotassium ethylenediaminetetra-acetic acid (EDTA) on platelet count and mean volume (MPV) was evaluated using two routine measurement systems, a Coulter S Plus (Phase 1) (S+) and a Technicon H6000 (H6000). In normal subjects (n = 29) MPV increased by 17% during 39 h storage in EDTA when measured by the S+. In contrast MPV decreased by 22% when measured by the H6000. MPV differences of up to 40% were observed between the two systems. Concomitant platelet counts, in both systems, changed by less … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

4
50
0

Year Published

1987
1987
2016
2016

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 64 publications
(54 citation statements)
references
References 17 publications
4
50
0
Order By: Relevance
“…21 First, platelets swell in a time-dependent fashion after blood sampling in EDTA. 19 The intention of conducting assays 24 to 48 hours after blood collection (when platelet swelling will have ceased 19,20,25 ) was intended to minimize such errors. However, a relatively large proportion was assayed outside of the proposed time window.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…21 First, platelets swell in a time-dependent fashion after blood sampling in EDTA. 19 The intention of conducting assays 24 to 48 hours after blood collection (when platelet swelling will have ceased 19,20,25 ) was intended to minimize such errors. However, a relatively large proportion was assayed outside of the proposed time window.…”
Section: Discussionmentioning
confidence: 99%
“…Second, the MPV measurements were made with the use of a different machine in each of the collaborating countries over a period of approximately 2 years. The quality control program identified machine-dependent differences between the average levels of MPV in the participating countries, especially between New Zealand (where the machine used an optical mechanism, in which measured MPV appears to decrease with swelling) and other countries (where machines used an electric impedance mechanism of assay, in which measured MPV tends to increase with swelling 25 ). Country stratification or a country dummy variable was used in all analyses, although such adjustment is unlikely to have controlled completely for the differences between the machines.…”
Section: Discussionmentioning
confidence: 99%
“…The data on the size of the platelets have interested physicians since the last three decades. [1][2][3] The recent advancements in automated hematology analyzers have added new dimensions to the platelet indices (mean platelet volume, plateletcrit, platelet distribution width and platelet -large cell ratio). While the data on red blood size (mean corpuscular volume) is universally acknowledged and even applied, the diagnostic and prognostic utility of platelet size (mean platelet volume) in thrombocytopenia is still in its nascent stage of exploration.…”
Section: Introductionmentioning
confidence: 99%
“…Conversely when MPV is measured by an optical light scatter system derived from the modal platelet size, the MPV decreases over time, possibly as a result of the dilution of cytoplasmic contents leading to a decrease in light scatter [6]. An equation allowing for the correction of MPV over time from venepuncture has been published [7], but this is not widely used because of the unpredictability in the time taken to equilibrate in individual samples.…”
mentioning
confidence: 99%