Abstract:By investigating the influence of initial electrons on dosimetric characteristics, reasonable incident electron parameters for the nominal 6 MV photon beam of the XHA600D accelerator are finally established, i.e., a 6 MeV monoenergetic electron beam with a radial intensity FWHM of 2.5 mm and an angular divergency of 0.15°. Based on reasonable initial parameters, Percentage Depth Doses (PDDs), Off-Axis Ratios (OARs), total scatter factors, beam qualities, and penumbra widths of both flatteningfilter (FF) and fl… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.