“…The WCD system is composed of 4 components: (1) a garment facilitating positioning of an electrode belt, (2) the electrode belt that incorporates 4 sensing (2 electrocardiograph channels) and 3 defibrillation nonadhesive electrodes, (3) charger, and (4) the monitor pack ( Figure 1). 3,4 The latter houses the rechargeable battery, the patient response button, and a touch screen that allows the patient to capture and transmit information about their rhythm, engage in a 6minute hall walk test, and enter information to ascertain quality-of-life (QoL) indices. Arrhythmia detection is technically challenging, as the 4 dry sensing electrodes are prone to skin motion artifact or loss of contact; this necessitates sophisticated compensatory algorithms to minimize false alarms and inappropriate therapy.…”