A commercial differential scanning calorimeter (DSC)
was adapted
to allow accurate isobaric heat capacity, c
p
, measurements of cryogenic, high-pressure liquids.
At (subcritical) temperatures between (108.15 and 258.15) K and pressures
between (1.1 and 6.35) MPa, the standard deviation in the measured c
p
values was 0.005·c
p
for methane, 0.01·c
p
for ethane, and 0.015·c
p
for propane, which is comparable
to both the scatter of c
p
data for these liquids measured using other techniques and with
the scatter of those data sets about the reference equation of state
(EOS) values. Three key modifications to the commercial DSC were required
to enable these accurate cryogenic, high-pressure liquid c
p
measurements. First, methods of loading
and removing the liquid from the calorimeter without moving the sample
cell were developed and tested with high boiling temperature liquids;
this modification improved the measurement repeatability. Second,
a ballast volume containing a high-pressure vapor phase at constant
temperature was connected to the DSC cell so that the liquid sample’s
thermal expansion did not cause significant changes in pressure. A
third modification was required because the boil-off vapor from the
liquid nitrogen used to cool the calorimeter resulted in a temperature
inversion, and hence convection, along the tubing connecting the DSC’s
sample cell to the ballast volume, that lead to an unstable calorimetric
signal at T > 130 K. The modifications to the
specialized
DSC were tested by measuring c
p
values for heptane, methylbenzene, and a heptane (1) + methylbenzene
(2) mixture with x
1 = 0.38 at atmospheric
pressure and temperatures of (228.15, 238.15, 303.15, and 313.15)
K. The measured values had relative deviations from those measured
adiabatically by Holzhauer and Ziegler (J. Phys. Chem. 1975, 79, 590–604) of less
than 0.006·c
p
, indicating
that the specialized DSC could be used for liquids and liquid mixtures
at conditions relevant to liquefied natural gas production.