Tighter precision target required for lactate testing in patients with lactic acidosis

Abstract: As such, evaluations of lactate measurements must use tighter allowable error limits.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
5
0

Year Published

2015
2015
2021
2021

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 8 publications
(5 citation statements)
references
References 5 publications
0
5
0
Order By: Relevance
“…Both POCT devices were evaluated for analytical imprecision by comparing the measured CV of internal quality controls with the CV provided by the company, the analytical imprecision and the Total Allowable Error (TEa) [18] , [19] . Lactate concentrations, measured with the StatStrip-Lactate, showed a CV of 5.5% (mean: 6.2 mmol/L), which is lower than the CV stated by the company and the target analytical imprecision (13.6%) for lactate, as recently published [16] . Lactate and pH measured with the iSTAT-CG4+ showed very low CV’s of 1.1% (mean: 6.94 mmol/L) and 0.1% (mean: 7.05), respectively.…”
Section: Resultsmentioning
confidence: 54%
See 3 more Smart Citations
“…Both POCT devices were evaluated for analytical imprecision by comparing the measured CV of internal quality controls with the CV provided by the company, the analytical imprecision and the Total Allowable Error (TEa) [18] , [19] . Lactate concentrations, measured with the StatStrip-Lactate, showed a CV of 5.5% (mean: 6.2 mmol/L), which is lower than the CV stated by the company and the target analytical imprecision (13.6%) for lactate, as recently published [16] . Lactate and pH measured with the iSTAT-CG4+ showed very low CV’s of 1.1% (mean: 6.94 mmol/L) and 0.1% (mean: 7.05), respectively.…”
Section: Resultsmentioning
confidence: 54%
“…Both POCT lactate devices studied showed good analytical performance and met the recently published tight precision target for testing patients with lactate acidosis [16] .…”
Section: Discussionmentioning
confidence: 57%
See 2 more Smart Citations
“…Table 1 summarizes published biological variation of CBC components expressed as a relative standard deviation [also known as coefficient of variation (CV b )]. We have been using a disruptive, data-mining approach to determine the biologic variation [4,5]. Our method does not require any additional laboratory testing.…”
Section: Introductionmentioning
confidence: 99%