The torsional performance is a major mechanical property of stent. A stent with good torsional performance is easy to deform along blood vessels without damaging the vascular wall to avoid in-stent restenosis (ISR). Therefore, this study aimed to study the effect of stent parameters on torsional performance. The effect of stent parameters on torsional performance was studied via finite element method (FEM). The twist metric (TM) and stress distribution of various stents were compared. The TM values of stents with I-, S-, M-, C-, and V-shaped linkers were 0.0190, 0.0191, 0.0184, 0.0141, and 0.0201[Formula: see text][Formula: see text], respectively. In addition, the TM value of the stent increased by 35.85 times when the number of linkers was increased from 2 to 8 and the stent was twisted at the same angular displacement in clockwise direction. The TM value of the stent with 1.13∘ tapering was 0.010 [Formula: see text], which was lower by 47.64% compared with that of cylindrical stent. Compared with the shape of the linker, the number of linkers had a more remarkable effect on torsional performance. Torsional performance was observably enhanced with the decrease in the number of linkers. Among the five stents with different linker shapes, the torsional performance of the stent with C-shaped linker was the best. Besides, the torsional performance of the tapered stent was better than that of the cylindrical stent. Moreover, the torsional performance increased by increasing the stent tapering. This work might provide insights into better stent design and clinical decisions.