2015
DOI: 10.1556/achrom.27.2015.2.1
|View full text |Cite
|
Sign up to set email alerts
|

Total error-based validation including the experimental design-based robustness evaluation of a stability-indicating method for the simultaneous quantification of hydrochlorothiazide and valsartan in tablet formulations

Abstract: Summary.A gradient reversed phase high-performance liquid chromatography (RP-HPLC) method with ultraviolet (UV) detection to analyze hydrochlorothiazide (HCT) and valsartan (VS) simultaneously in a tablet formulation during forced degradation studies was developed. This method was validated using a novel approach, namely, the accuracy profile or total errors approach. The robustness of the method was evaluated using a Plackett-Burman design for eight factors. The algorithm of Dong was applied to determine the … Show more