TU‐C‐304A‐03: Stylized MIRD Phantoms Should Be Replaced by Anatomically Realistic Phantoms: Discrepancies In Red Bone Marrow Doses From CT Scans

Abstract: Purpose: To test the hypothesis that the stylized MIRD phantoms would cause significant error in the estimated red bone marrow (RBM) dose from CT scans in comparison with anatomically realistic phantoms. Method and Materials: The MC model of the CT scanner include the source geometry, movement, source energy spectrum, bow‐tie filter, as well and the beam shape. MCNPX 2.5.0 was used to simulate the RBM dose from various CT scanning procedures. To calculate the absorbed dose to the RBM as a function of photon fl… Show more