2006
DOI: 10.1093/rpd/ncm158
|View full text |Cite
|
Sign up to set email alerts
|

Uncertainties in determination of 131I activity in thyroid gland

Abstract: The values of the iodine 131I uptake determined from the measurements with a NaI scintillation counter might be significantly influenced by the thyroid position in the human neck. It is shown that the ratio of the counting rate in the energy peak of 364 keV to the counting rate in the Compton scattering band can be used for the determination of the effective depth of the thyroid. The uncertainties of the standard method are discussed basing on results of calibration measurements. The investigations of 95 patie… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
3
0
1

Year Published

2007
2007
2023
2023

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 9 publications
(4 citation statements)
references
References 0 publications
0
3
0
1
Order By: Relevance
“…The influence of other parameters such as vertical and horizontal shifts of the source from the detector axis are described elsewhere [1,3].…”
Section: Resultsmentioning
confidence: 99%
See 1 more Smart Citation
“…The influence of other parameters such as vertical and horizontal shifts of the source from the detector axis are described elsewhere [1,3].…”
Section: Resultsmentioning
confidence: 99%
“…The most important application are measurements of iodine uptake by patients with thyroid diseases. The P/C ratio can be used for selecting a group of patients for whom the anatomical position of the thyroid or the spatial distribution of the iodine accumulated is very much different from the standard distribution assumed during the calibration of the counters [1][2][3]. Standard measurements in this case give incorrect results.…”
Section: Resultsmentioning
confidence: 99%
“…pomiarów skażeń tarczycy jodem promieniotwórczym za pomocą licznika promieniowania tarczycy. Kalibrację licznika można wykonać za pomocą fantomów pozwalających na symulowanie różnego położenia tarczycy w szyi człowieka [4,5].…”
Section: Wnioskiunclassified
“…size, shape, etc.) and the position of each organ in human body [ 1 3 ]. The thyroid was chosen as an object of measurements.…”
Section: Introductionmentioning
confidence: 99%