2015
DOI: 10.5755/j01.eee.21.1.6834
|View full text |Cite
|
Sign up to set email alerts
|

Unconventional Double-Current Circuit for Deflection and Temperature Simultaneous Measurement

Abstract: In this paper some interesting and novel features of a four-terminal (4T) network are presented. A single DC current source (J) is switched over and connected in turns to opposite arms of the four-element bridge circuit. This twooutput circuit with two voltage outputs is called a double current circuit (2x1J). The output voltages are differently dependent on the arm resistance increments and their values are given in absolute and relative units. An original application with two sensors acting as strain gauges … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
7
0

Year Published

2016
2016
2022
2022

Publication Types

Select...
5
3

Relationship

2
6

Authors

Journals

citations
Cited by 12 publications
(9 citation statements)
references
References 8 publications
0
7
0
Order By: Relevance
“…The results are given in Tab In the case of the k2 coefficient it can be assumed that the X value, for which the calibration is carried out, is not important (the change in this case is only 0.05%). The following fixed values were assumed in order to conduct the calculations: ad=0.00832 mm -1 [2], W=100, α=0.00004 °C -1 , R0=121.116 Ω.…”
Section: Metrological Estimation Of the Results Differencesmentioning
confidence: 99%
See 1 more Smart Citation
“…The results are given in Tab In the case of the k2 coefficient it can be assumed that the X value, for which the calibration is carried out, is not important (the change in this case is only 0.05%). The following fixed values were assumed in order to conduct the calculations: ad=0.00832 mm -1 [2], W=100, α=0.00004 °C -1 , R0=121.116 Ω.…”
Section: Metrological Estimation Of the Results Differencesmentioning
confidence: 99%
“…The output voltages of this direct current bridge are functions of differences and sums of the resistance increases of two foil strain gauges attached to a cantilever beam. Such circuit can be an alternative for system solutions in the case when deflection and temperature need to be recorded in the same point of the measured element (Idzkowski et al, 2015;Cappa et al, 2001;Paker, 1993). Also other solutions can be treated as differential measurement circuits, e.g.…”
Section: Introductionmentioning
confidence: 99%
“…The graphs of relative non-linearity coefficients (12), (13) are presented in Figs. 2-5 (for J = 100 mA and R 0 = 100 Ω).…”
Section: Introductionmentioning
confidence: 99%
“…In order to test its metrological qualities, an attempt to physically realize the idea was taken. It was done through modifying the way of supply by introducing one, switchable current source [13].…”
Section: Introductionmentioning
confidence: 99%
“…The other one has an unconventional supplying: the 4T circuit is supplied by two equal current sources J connected to opposite arms -2J which is realized by switching over two equal sources between these arms -2×2J. In some cases only one current supply can be used -2×1J [4].…”
Section: Introductionmentioning
confidence: 99%