Understanding Sensitivity and Cross-Reactivity of Xylazine Lateral Flow Immunoassay Test Strips for Drug Checking Applications

Edward Sisco,
Danielle F. Nestadt,
Madeline B. Bloom
et al.

Abstract: The continued prevalence of xylazine in the illicit drug market has necessitated the development of quick and simple methods for identification of adulterated materials. One method that could be employed is lateral flow immunoassays (also known as test strips), which are similar to those frequently used to determine the presence of fentanyl. In this work, we explored the use of xylazine test strips (XTS) for the detection of xylazine in drug residue samples and investigated their sensitivity and cross-reactivi… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

0
4
0

Year Published

2023
2023
2023
2023

Publication Types

Select...
1

Relationship

0
1

Authors

Journals

citations
Cited by 1 publication
(4 citation statements)
references
References 19 publications
0
4
0
Order By: Relevance
“…20 Early results from the John Hopkins Bloomberg School of Public Health regarding the use of the xylazine test strips for drug checking show the overall sensitivity of the test strips to be good and cross-reactivity to be limited to a single compound -lidocaine. 21 The xylazine test strips were found to detect xylazine at concentrations as low as 0.5 mcg/mL in solution and consistently produced a positive result at concentrations as low as 2.5 mcg/mL. 21 When the xylazine test strips were used to detect xylazine in realworld drug residue samples, xylazine was consistently detected in samples with a concentration greater than 2 mcg/mL; there were mixed results when the concentration ranged between 0.1 mcg/mL and 2 mcg/ mL.…”
Section: What Is the Evidence?mentioning
confidence: 99%
See 3 more Smart Citations
“…20 Early results from the John Hopkins Bloomberg School of Public Health regarding the use of the xylazine test strips for drug checking show the overall sensitivity of the test strips to be good and cross-reactivity to be limited to a single compound -lidocaine. 21 The xylazine test strips were found to detect xylazine at concentrations as low as 0.5 mcg/mL in solution and consistently produced a positive result at concentrations as low as 2.5 mcg/mL. 21 When the xylazine test strips were used to detect xylazine in realworld drug residue samples, xylazine was consistently detected in samples with a concentration greater than 2 mcg/mL; there were mixed results when the concentration ranged between 0.1 mcg/mL and 2 mcg/ mL.…”
Section: What Is the Evidence?mentioning
confidence: 99%
“…21 The xylazine test strips were found to detect xylazine at concentrations as low as 0.5 mcg/mL in solution and consistently produced a positive result at concentrations as low as 2.5 mcg/mL. 21 When the xylazine test strips were used to detect xylazine in realworld drug residue samples, xylazine was consistently detected in samples with a concentration greater than 2 mcg/mL; there were mixed results when the concentration ranged between 0.1 mcg/mL and 2 mcg/ mL. 21 These findings about the limitations of these test strips in the smallest concentrations are pertinent to the overall effectiveness of the strips to detect xylazine in the unregulated drug supply.…”
Section: What Is the Evidence?mentioning
confidence: 99%
See 2 more Smart Citations