Using Collision Cross Section Distributions to Assess the Distribution of Collision Cross Section Values

Abstract: In this study we explore the use of collision cross section distributions to allow comparability of IM-MS data for proteins on different instruments. We present measurements on seven standard proteins across three IM-MS configurations, namely an Agilent 6560 IM QToF, a Waters Synapt G2 possessing a TWIMS cell and a modified Synapt G2 possessing an RF confining linear field drift cell. Mobility measurements were taken using both He and N<sub>2</sub> as the drift gases. To aid comparability across i… Show more

Help me understand this report
View published versions

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
10
0

Year Published

2021
2021
2023
2023

Publication Types

Select...
2
2
1

Relationship

1
4

Authors

Journals

citations
Cited by 5 publications
(10 citation statements)
references
References 0 publications
0
10
0
Order By: Relevance
“…To gain more insightful information on the structure of the molecules, the ATD can be converted to a distribution of the CCS values measured. 10,182 As mentioned in detail (sections 3.1, 3.2) DT IM-MS does not require calibration with known proteins' CCS values (Figure 2a). Hence, in this section, we will only consider the case of TWIMS where calibration is mandatory to convert the ATD to CCS distributions using known calibrants (Figure 2b).…”
Section: Obtaining Ccs Values From Experimental Datamentioning
confidence: 99%
See 2 more Smart Citations
“…To gain more insightful information on the structure of the molecules, the ATD can be converted to a distribution of the CCS values measured. 10,182 As mentioned in detail (sections 3.1, 3.2) DT IM-MS does not require calibration with known proteins' CCS values (Figure 2a). Hence, in this section, we will only consider the case of TWIMS where calibration is mandatory to convert the ATD to CCS distributions using known calibrants (Figure 2b).…”
Section: Obtaining Ccs Values From Experimental Datamentioning
confidence: 99%
“…TWIMS separation follows the same concepts as DTIMS where; expanded conformers surfing the wave, have a higher surface area available for collisions with the inert gases in the stack resulting in delayed arrivals to the detector. However, unlike DTIMS, TWIMS does not provide direct information on the CCS of the analytes from first principles 10,182 but instead requires calibration using standards of known CCS values. 10,183,184 More on the calibration approach is discussed in section 3.5.…”
Section: Traveling Wave Ion Mobility Spectrometry (Twims)mentioning
confidence: 99%
See 1 more Smart Citation
“…27,28 Several commercially available IM-based devices use variations in the applied electric field and stationary state of the buffer gas. 29,30 For instance, the drift tube IMS (DTIMS) employs a fixed voltage drop along the length of the drift tube to generate a relatively weak uniform electric field under which ions move through the drift tube. Unlike the other commercial IM−MS instruments, DTIMS can provide direct CCS measurements without calibration.…”
Section: ■ Introductionmentioning
confidence: 99%
“…For CCS, despite the still scarce availability of online databases, several CCS predictive models have been proliferating (Bijlsma et al, 2019;Colby et al, 2019;Gonzales et al, 2016;Mollerup et al, 2018;Plante et al, 2019;Ross et al, 2020;Zhou et al, 2017Zhou et al, , 2016. Recent studies over the reproducibility of CCS measurement over different instruments highlighted that CCS predictive models can be easily incorporated into other laboratories provided that certain conditions are fulfilled such as IMS technique and/or CCS calibrants used (France et al, 2020;Hinnenkamp et al, 2018;Righetti et al, 2020;Stow et al, 2017).…”
Section: In-silico Prediction Toolsmentioning
confidence: 99%