Variability in Infusion Pressure and Continuous Flow Rate Delivered from Pressurized Bag Pump Flush Systems

Abstract: Using 10 different infusion bag pressure pumps, indicated manometer pressures were compared with measured infusion pressures proximal to the flow-regulating device in an in vitro experiment. Flow rates delivered through the flow-regulating device were gravimetrically measured at different monitored pressure levels. Significant differences were found between manometer and measured infusion pressures among the tested pressure bag pumps (e.g. 500 ml bag volume pressurized to 300 mmHg manometer pressure: 219.6±7.8… Show more