Abstract:Using 10 different infusion bag pressure pumps, indicated manometer pressures were compared with measured infusion pressures proximal to the flow-regulating device in an in vitro experiment. Flow rates delivered through the flow-regulating device were gravimetrically measured at different monitored pressure levels. Significant differences were found between manometer and measured infusion pressures among the tested pressure bag pumps (e.g. 500 ml bag volume pressurized to 300 mmHg manometer pressure: 219.6±7.8… Show more
P Pu ur rp po os se e: : To evaluate the risk of retrograde embolization of flush solution in neonates and infants with routinely used electronic syringe pumps and infusion bag pump flush systems.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.