Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla

Abstract: Purpose: To evaluate MRI issues at 3-Tesla for a ventricular assist device (VAD). Methods:The AB5000 Ventricle with a prototype Nitinol wire-reinforced In-Flow Cannula and Out-Flow Cannula attached (Abiomed, Inc., Danvers, MA) was evaluated for magnetic field interactions, heating, and artifacts at 3-Tesla. MRI-related heating was assessed with the device in a gelled-saline-filled, head/torso phantom using a transmit/received RF body coil while performing MRI at a whole body averaged SAR of 3-W/kg for 15-min. … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
5

Citation Types

0
15
0

Year Published

2008
2008
2012
2012

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 6 publications
(15 citation statements)
references
References 12 publications
0
15
0
Order By: Relevance
“…17,[20][21][22][23] Translational Attraction. To evaluate translational attraction for each aneurysm clip, the deflection-angle technique was used according to a previously described methodology.…”
Section: Magnetic Field Interactionsmentioning
confidence: 99%
See 4 more Smart Citations
“…17,[20][21][22][23] Translational Attraction. To evaluate translational attraction for each aneurysm clip, the deflection-angle technique was used according to a previously described methodology.…”
Section: Magnetic Field Interactionsmentioning
confidence: 99%
“…The test fixture incorporated a protractor with 1°graduated markings. 17,[20][21][22][23] The test sample was suspended on the apparatus by a lightweight string (20 cm in length; weight, Ͻ1% of the weight of each clip), which was fixed at the 0°i ndicator of the protractor. Deflection angles for each aneurysm clip were assessed at the point of the highest spatial magnetic gradient for the 3T MR system.…”
Section: Magnetic Field Interactionsmentioning
confidence: 99%
See 3 more Smart Citations