Verification of dynamic and segmental IMRT delivery by dynamic log file analysis

Abstract: A program has been developed to evaluate the delivered fluence of step‐and‐shoot segmental and sliding window dynamic multileaf collimator (MLC) fields. To automate these checks, a number of tools have been developed using data available from the dynamic log files that can be created each time a dynamic delivery occurs. Experiments were performed with a Varian 2100EX with a 120 leaf MLC equipped with dynamic capabilities. A dynamic leaf sequence is delivered and measured with film or an amorphous silicon image… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
5

Citation Types

1
65
0

Year Published

2012
2012
2020
2020

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 77 publications
(66 citation statements)
references
References 10 publications
1
65
0
Order By: Relevance
“…One approach to MLC QA is to analyze the machine's logfile to perform MLC QA,2, 3 and this approach can be used with data from a dynamic delivery (MLC position, Gantry Angle, and Monitor Unit) with higher sampling rate (10–50 ms). In addition, logfile QA can be performed using three‐dimensional dose reconstruction based on patient geometry, and several recent reports have described this technique 4, 5, 6…”
Section: Introductionmentioning
confidence: 99%
“…One approach to MLC QA is to analyze the machine's logfile to perform MLC QA,2, 3 and this approach can be used with data from a dynamic delivery (MLC position, Gantry Angle, and Monitor Unit) with higher sampling rate (10–50 ms). In addition, logfile QA can be performed using three‐dimensional dose reconstruction based on patient geometry, and several recent reports have described this technique 4, 5, 6…”
Section: Introductionmentioning
confidence: 99%
“…The log files analysis will also report delivered dose errors due to dropped segments, leaf over/under shoot, faulty leaf motors, tongue‐and‐grove effect, rounded leaf ends, and communication delays between the controller and the workstation. In addition, they also tell us if the MLC gain calibration is set appropriately 8 , 11 …”
Section: Introductionmentioning
confidence: 99%
“…Methods for analysis and presentation of machine log files have been previously reported for linear accelerator‐(linac‐) based external beam radiation therapy (EBRT) treatments 9 , 10 , 11 , 12 , 13 . The MLC dynamic log files are available on Varian linac treatment machines (14) and have been utilized at the authors' institution and by other groups to validate the accuracy and completeness of the treatment deliveries 15 , 16 , 17 , 18 , 19 , 20 , 21 .…”
Section: Introductionmentioning
confidence: 99%