Verifikasi Metode Uji COD secara Spektrofotometri UV-Vis untuk Low Concentration dan High Concentration

Abstract: Telah dilakukan verifikasi metode uji chemical oxygen demand (COD) dengan metode spektrometri UV-Vis secara refluks tertutup. Verifikasi metode ini dilakukan untuk mengkonfirmasi metode SNI 6989.2:2019 untuk nilai COD low concentration dan high concentration. Parameter verifikasi metode meliputi penentuan lineritas, method detection limit, limit of quantification, presisi, dan akurasi. Berdasarkan hasil penelitian menunjukkan bahwa metode penentuan COD untuk low concentration dan high concentration memiliki li… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2

Citation Types

0
2
0

Year Published

2023
2023
2023
2023

Publication Types

Select...
1

Relationship

0
1

Authors

Journals

citations
Cited by 1 publication
(2 citation statements)
references
References 8 publications
0
2
0
Order By: Relevance
“…Figure 1 shows testing on the high concentration range, which is based on chromium (III) measurement from the reduction of potassium dichromate by organic compounds in the test sample [6]. Based on the calculation of the linearity value, the regression equation is Y=0,0003x +0,0073, and the correlation coefficient value is 0.9984.…”
Section: Figure 1 High Standard Solution Calibration Curvementioning
confidence: 99%
See 1 more Smart Citation
“…Figure 1 shows testing on the high concentration range, which is based on chromium (III) measurement from the reduction of potassium dichromate by organic compounds in the test sample [6]. Based on the calculation of the linearity value, the regression equation is Y=0,0003x +0,0073, and the correlation coefficient value is 0.9984.…”
Section: Figure 1 High Standard Solution Calibration Curvementioning
confidence: 99%
“…Figure 2 shows the test results in the low concentration range based on chromium (VI) measurement from the rest of the potassium dichromate used in the oxidation reaction of organic compounds [6]. Measurements were made at a wavelength of 420 nm, and the occurrence of the inverse ratio between standard concentration and absorbance so that the resulting curve decreases.…”
Section: Figure 2 Low Standard Solution Calibration Curvementioning
confidence: 99%