2002
DOI: 10.1067/mpr.2002.128099
|View full text |Cite
|
Sign up to set email alerts
|

Wave analysis of implant screw loosening using an air cylindrical cyclic loading device

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

1
30
0
8

Year Published

2005
2005
2019
2019

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 35 publications
(39 citation statements)
references
References 15 publications
1
30
0
8
Order By: Relevance
“…Factors that influence preload are the geometry of the screw (thread pitch) and the mechanical properties of metal, including texture, surface finishing, surface settlement, and degree of lubrication. [591624252627]…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…Factors that influence preload are the geometry of the screw (thread pitch) and the mechanical properties of metal, including texture, surface finishing, surface settlement, and degree of lubrication. [591624252627]…”
Section: Discussionmentioning
confidence: 99%
“…However, published studies have used different methodologies for the applied axial load, frequency, and number of cycles. [52627] The frequency of cycles is reported in the literature as ranging from 1 to 19 Hz. [27] It was clarified that in a day, an individual typically performs three episodes of chewing lasting 15 min, with a frequency of 60 cycles per minute (1 Hz); this generates 2700 chewing cycles per day, which equals 1 million cycles per annum.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…As part of a study to clarify the features and advantages of internal anti-rotation configurations, the present study aimed to investigate the effect of cyclic loading on abutment screw loosening in internal and external hexagon implants. To date, only a few reports are available on abutment screw joint stability in external hexagon implant systems under eccentric cyclic loading [28][29][30][31] . Moreover, little or no data exist on screw loosening in internal hexagon implant systems under eccentric cyclic loading 32) .…”
Section: Introductionmentioning
confidence: 99%
“…1 Although a low incidence of screw loosening or fracture has been reported, the resolution of such problems is time consuming, sometimes requiring multiple appointments. [2][3][4] The depth and the possibility of having the screw fragment wedged in a damaged area of the implant internal threads are factors that may influence the complexity and time involved in the retrieval attempt. 4 A simplified technique for the removal of fractured screw fragments is described.…”
mentioning
confidence: 99%