In radiography, one of the best methods to eliminate image-degrading
scatter radiation is the use of anti-scatter grids. However, with
high-resolution dynamic imaging detectors, stationary anti-scatter grids can
leave grid-line shadows and moiré patterns on the image, depending upon
the line density of the grid and the sampling frequency of the x-ray detector.
Such artifacts degrade the image quality and may mask small but important
details such as small vessels and interventional device features. Appearance of
these artifacts becomes increasingly severe as the detector spatial resolution
is improved. We have previously demonstrated that, to remove these artifacts by
dividing out a reference grid image, one must first subtract the residual
scatter that penetrates the grid; however, for objects with anatomic structure,
scatter varies throughout the FOV and a spatially differing amount of scatter
must be subtracted.
In this study, a standard stationary Smit-Rontgen X-ray grid (line
density - 70 lines/cm, grid ratio - 13:1) was used with a high-resolution CMOS
detector, the Dexela 1207 (pixel size - 75 micron) to image anthropomorphic head
phantoms. For a 15 × 15cm FOV, scatter profiles of the anthropomorphic
head phantoms were estimated then iteratively modified to minimize the
structured noise due to the varying grid-line artifacts across the FOV.
Images of the anthropomorphic head phantoms taken with the grid, before
and after the corrections, were compared demonstrating almost total elimination
of the artifact over the full FOV. Hence, with proper computational tools,
anti-scatter grid artifacts can be corrected, even during dynamic sequences.