XEN®-63 Compared to XEN®-45 Gel Stents to Reduce Intraocular Pressure in Glaucoma

Charlotte Evers,
Daniel Böhringer,
Sara Kallee
et al.

Abstract: The XEN® gel stent reduces intraocular pressure (IOP) in glaucoma. XEN®-45 is widely used while the newer XEN®-63 has a larger lumen that may lower IOP more. We retrospectively compared the first 15 XEN®-63 cases to 15 matched XEN®-45 controls. With a preoperative IOP of 18.1 ± 3.9 mmHg (mean ± SD) and a final IOP of 9.2 ± 4.2 mmHg, XEN®-63 implantation resulted in an IOP reduction of 48.0 ± 23.3%. Similarly, with a preoperative IOP of 18.3 ± 4.5 mmHg and a final IOP of 10.5 ± 2.3 mmHg, XEN®-45 implantation re… Show more