We have developed a high-pressure electroluminescence (EL) detector consisting of a cathode, 5 drift rings, and an EL-generating structure which defines a sensitive volume of ∅5cm x 5cm. The EL-generating structure consists of two parallel-plate, chemically etched grids and a high-pressure optical window which is optically coupled to an external photomultiplier tube. Ionizing radiation that is absorbed in the sensitive volume generates electrons, which drift into the EL region and produce an EL flash. The detector was filled with 20bar Xe gas that was highly purified (>1 msec electron-life time) using a spark purification technique. In order to evaluate the potential for using an EL detector under adverse conditions, we disturbed the detector using a 10 W electric engraver working at 60 Hz. The action of the engraver had practically no effect on spectra acquired from an 241 Am source. I. ChannelsCounts 0-7803-8257
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.