The CliniMACS CD34+ selection device was used for positive selection of apheresis products for autologous transplantation from 10 patients with malignant diseases and for allogeneic transplantation from 26 healthy donors. A total of 71 separations were performed. In 1 allogeneic donor, CD34+ progenitors were also isolated from bone marrow. Between 0.27 and 8.9 x 10(10) nucleated cells (median 4.9 x 10(10)) containing 0.09%-10.8% (median 0.67%) CD34+ progenitor cells were separated. After separation, a median number of 227 x 10(6) mononuclear cells (MNC) (51-524) were recovered, with a median viability of 99% (22%-100%) and a median purity of 97.0% (68.3%-99.7%) CD34+ cells. Depletion of T cells was extensive, with a median of 0.04% residual CD3+ cells (range <0.01%-0.92%). Residual CD19+ cells were between <0.01% and 17%, including CD34+CD19+ cells. Recovery of CD34+ cells was calculated according to the ISHAGE guidelines and ranged from 24% to 105% (median 71%). We conclude that with the CliniMACS device CD34+ cells with high purity and recovery can be isolated with concomitant effective T cell depletion in the allogeneic setting and with a high purging efficacy in the autologous setting.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.