Two strategies have been developed for Diuron determination by FTIR spectrometry, an off-line extraction and stopped-flow determination and a fully mechanized procedure, based on the on-line extraction of Diuron and FIA-FTIR measurement of the extracts. The aforementioned procedures have been compared with a reference chromatographic method. The off-line FTIR spectra were obtained at a nominal resolution of 4 cm(-1) from 4000 to 900 cm(-1) by accumulating 25 scans. Diuron was determined using peak height measurements at 1582 cm(-1) corrected using a baseline defined between 1562 and 1614 cm(-1). The waste generation of the off-line procedure was 3.4 mL chloroform for each sample, and the method provided a LOD of 40 microg g(-1), corresponding to 0.8% (w/w) Diuron in the original sample. The fully mechanized FIA method provided a LOD of 35 microg g(-1), which corresponds to 0.7% (w/w) in the solid sample and a maximum sampling frequency of the whole procedure of 30 h(-1), with a waste generation of 9.3 mL per sample, taking into account the volume of CHCl(3) required for sample dissolution and that need as a carrier. All those methods consume less organic solvent than a HPLC method, which involves the use of 39 mL of acetonitrile per sample and a sampling frequency of 12 h(-1).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.