The Berthold LB950 Automatic Luminescence Analyser was used to estimate bacterial adenosine triphosphate in urine. The system provided a rapid (15 min) and fully automated screening test for bacteriuria at the 10(5) CFU/ml level. Bioluminescence results for 1040 urines were compared with viable counts using two reference culture methods and frequency distributions of bacterial counts and adenosine triphosphate levels were calculated. With a specificity of 79% the automated method showed a sensitivity of 84% using a pour plate reference count and 91% using a standard loop reference count. When contaminated urines were excluded the sensitivity improved to 98%. The automated bioluminescence test, though expensive, was shown to work well with good quality specimens.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.