This paper studies the effect of contact stress on friction and wear of ultra-high molecular weight polyethylene (UHMWPE) acetabular cups by means of friction and wear joint simulator testing under serum lubrication. For a given applied load, increasing the contact stress by increasing the ball/socket radial clearance decreased both the coefficient of friction and the wear rate. Friction and wear were highly correlated. The dependence of friction on contact stress for the UHMWPE socket under serum lubrication was similar to that of semi-crystalline polymers under dry sliding. This finding indicates the occurrence of partial dry contact at asperity levels for the metal-polyethylene ball-in-socket joint under serum lubrication.
There have been increasing concerns about the structural fatigue resistance of crosslinked UHMWPE devices due to deterioration of certain mechanical properties. However, due to the lack of clear correlation between specific mechanical properties and clinical performance, these concerns remain theoretical. In order to evaluate the potential benefits and risks of various crosslinked polyethylene materials for hip and knee bearing applications, two clinically relevant worst-case scenarios were simulated on functional devices. In the first worst-case scenario, cemented all-poly patellar components were tested under simulated stair-climbing conditions with rotational misalignment. In the second worst-case scenario, metal-backed thin acetabular liners were tested in a hip joint simulator under rim-loading conditions. Various types of crosslinked UHMWPE were prepared according to published process descriptions of commercial materials. While significant levels of volumetric wear reduction were confirmed by both the patellar and hip simulator tests, mixed results were obtained on the structural integrity of the devices. The latter was more significantly affected by the post-irradiation thermal treatment history than by the total dose of irradiation. Re-melting following irradiation led to catastrophic fractures of both rim-loaded liners and rotational-malaligned patellar pegs. The key mechanical property that was positively identified to correlate with the structural fatigue performance of crosslinked polyethylene materials was the ultimate tensile strength, whereas tensile elongation within 250% and 400% range had no effect on structural integrity. However, the results presented in this study should not be misconstrued with respect to the potential clinical performance of irradiation crosslinked and re-melted polyethylene liners with proper designs. In fact, the true outcome of all the crosslinked materials can only be revealed by long-term clinical follow-up.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.