An increasing number of studies have shown that post-mastectomy radiotherapy presents benefits associated with the patients survival and a significant fraction of the treated patients makes use of tissue expanders for breast reconstruction. Some models of tissue expanders have a magnetic disk on their surface that constitutes heterogeneity in the radiation field, which can affect the dose distribution during the radiotherapy treatment. In this study, the influence of a metallic heterogeneity positioned in a breast tissue expander was evaluated by means of Monte Carlo simulations using the MCNPX code and using Eclipse treatment planning system. Deposited energy values were calculated in structures which have clinical importance for the treatment. Additionally, the effect in the absorbed energy due to backscattering and attenuation of the incident beam caused by the heterogeneity, as well as due to the expansion of the prosthesis, was evaluated in target structures for a 6 MV photon beam by simulations. The dose distributions for a breast treatment were calculated using a convolution/superposition algorithm from the Eclipse treatment planning system. When compared with the smallest breast expander volume, underdosage of 7% was found for the largest volume of breast implant, in the case of frontal irradiation of the chest wall, by Monte Carlo simulations. No significant changes were found in dose distributions for the presence of the heterogeneity during the treatment planning of irradiation with an opposed pair of beams. Even considering the limitation of the treatment planning system, the results obtained with its use confirm those ones found by Monte Carlo simulations for a tangent beam irradiation. The presence of a heterogeneity didńt alters the dose distributions on treatment structures. The underdosage of 7% observed with Monte Carlo simulations were found for irradiation at 0°, not used frequently in a clinical routine.
The combination of radiotherapy treatments and breast reconstruction, using temporary tissue expanders, generates several concerns due to the presence of a magnetic valve inside the radiation field. The objective of this work is to evaluate a radiotherapy treatment planning for a patient using a tissue expander. Isodose curve maps, obtained using radiochromic films, were compared to the ones calculated with two different dose calculation algorithms of the Eclipse radiotherapy Treatment Planning System (TPS), considering the presence or absence of the heterogeneity. The TPS calculation considering the presence of the heterogeneity shows changes around 5% in the isodose curves when they were compared with the calculation without heterogeneity correction. This calculation did not take in account the real density value of the heterogeneity. This limitation was quantified to be around 10% in comparison with the TPS calculation and experimental measurements using the radiochromic film. These results show that the magnetic valve should be taken in account in dose calculations of the TPS. With respect to the AAA and Pencil Beam Convolution algorithms, when the calculation is compared with the real distribution, AAA presents a distribution more similar to experimental dose distribution.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.