Objectives
We aimed to compare bare‐metal stents (BMS), durable‐polymer everolimus‐eluting stents (DP‐EES), and abluminal biodegradable‐polymer sirolimus‐eluting stents (ABP‐SES) in the bifurcation model setup.
Background
The mechanism of thrombogenicity, which differs among second‐generation stents implanted using double‐kissing (DK) crush or culotte stenting techniques, remains unclear. We have shown previously that setting up a porcine arteriovenous shunt model is feasible and useful to assess thrombogenicity at vessel bifurcation points.
Methods
Six porcine shunt models were prepared for the comparison between DK crush and culotte stenting techniques using BMS, DP‐EES, and ABP‐SES. Intracoronary imaging with high‐resolution optical coherence tomography (OCT) was performed to evaluate the thrombogenicity in different stent types in the bifurcation stenting model and was evaluated by a core lab.
Results
Culotte stenting demonstrated more thrombogenicity at the proximal main branch (MB) with DP‐EES, side branch (SB) with BMS, and the bifurcation site irrespective of the stent type, while DK crush technique exhibited thrombogenicity only at SB with BMS and ABP‐SES. OCT analysis revealed malapposition of DP‐EES in the proximal MB with culotte stenting. Stent expansion was generally larger in ABP‐SES than BMS and DP‐EES.
Conclusions
The study provides hypothesis‐generating findings in distinct thrombogenicity of bifurcation stenting with DP‐ or ABP‐coated drug‐eluting stents.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.