Magnetic Resonance-Electrical Properties Tomography (MR-EPT) is an imaging modality that maps the spatial distribution of the electrical conductivity and permittivity using standard MRI systems. The presence of a body within the scanner alters the RF field, and by mapping these alterations it is possible to recover the electrical properties. The field is time-harmonic, and can be described by the Helmholtz equation. Approximations to this equation have been previously used to estimate conductivity and permittivity in terms of first or second derivatives of RF field data. Using these same approximations, an inverse approach to solving the MR-EPT problem is presented here that leverages a forward model for describing the magnitude and phase of the field within the imaging domain, and a fitting approach for estimating the electrical properties distribution. The advantages of this approach are that 1) differentiation of the measured data is not required, thus reducing noise sensitivity, and 2) different regularization schemes can be adopted, depending on prior knowledge of the distribution of conductivity or permittivity, leading to improved image quality. To demonstrate the developed approach, both Quadratic (QR) and Total Variation (TV) regularization methods were implemented and evaluated through numerical simulation and experimentally acquired data. The proposed inverse approach to MR-EPT reconstruction correctly identifies contrasts and accurately reconstructs the geometry in both simulations and experiments. The TV regularized scheme reconstructs sharp spatial transitions, which are difficult to reconstruct with other, more traditional approaches.
A rotational Electrical Impedance Tomography (rEIT) methodology is described and shown to produce spatially accurate absolute reconstructions with improved image contrast and an improved ability to distinguish closely spaced inclusions compared to traditional EIT on data recorded from cylindrical and breast-shaped tanks. Rotations of the tank without altering the interior conductivity distribution are used to produce the rEIT data. Quantitatively, rEIT was able to distinguish two inclusions that were 1.5 cm closer together than traditional EIT could achieve for inclusions placed 2 to 3 cm from the center for the cylindrical tank, and rEIT was able to distinguish two tumor-like inclusions where traditional EIT could not reliably do so.Mathematical analysis showed that rEIT improves the number of stable singular vectors by up to 4.2 and 4.7 times than that of traditional EIT for the cylindrical and breast-shaped tanks, respectively, which is an indication of improved resolution. Direct investigations into measurements revealed minimum rotation angles that should yield data uncorrupted by noise. Two inverse approaches (one that inverts then fuses the data (I/DF) and one that fuses the data then inverts (DF/I)) and two mesh modeling approaches were considered. It was found that DF/I produces far better results compared to I/DF and a rotated-mesh approach produces further improvements. The ability to obtain improved absolute reconstructions using rEIT on a practical clinical scenario (breast-shaped tank experiment) is an important step towards using rEIT to improve previous EIT results in medical applications.
Prostate cancer (PCa) recurrences are often predicted by assessing the status of surgical margins (SM) – positive surgical margins (PSM) increase the chances of biochemical recurrence by 2–4 times which may lead to PCa recurrence. To this end, an electrical impedance acquisition system with a microendoscopic probe was employed in an ex-vivo study of human prostates. This system measures the tissue bioimpedance over a range of frequencies (1 kHz to 1MHz), and computes a number of Composite Impedance Metrics (CIM). A classifier trained using CIM data can be used to classify tissue as benign or cancerous. The system was used to collect the impedance spectra from 14 excised prostates, which were obtained from men undergoing radical prostatectomy, for a total of 23 cancerous and 53 benign measurements. The data revealed statistically significant (p<0.05) differences in the impedance properties of the benign and tumorous tissues, and among the measurements taken on the apical, base, and lateral surface of the prostate. Further, in the leave-one-patient-out cross validation, a maximum predictive accuracy of 90.79% was achieved by combining high frequency CIM phase data to train a support vector machine classifier with a radial basis function kernel. The observations are consistent with the physiology and morphology of benign and malignant prostate tissue. CIMs were found to be an effective tool in distinguishing benign from cancerous tissues.
The ability to obtain accurate conductivity values of a suspicious lesion (>1.8 cm) detected from another modality (e.g. AWB-US) could significantly reduce false positives and result in a clinically important technology.
A novel regularization technique is developed for end-fired microendoscopic electrical impedance tomography using the dual-mesh method. The new regularization technique coupled with appropriate forward modeling and inverse mesh design is shown to produce dramatically improved reconstructions over previous methods. 3D absolute and difference reconstructions from measured saline tank and ex vivo adipose and muscle tissue experiments are used to validate the approach. The ex vivo experiments are used as a surrogate for prostate tissue, which is the primary clinical application for the probe. Inclusion center of mass errors were less than 0.47 mm for tank experiments with inclusion depths and radial offsets ranging less than 3 mm and 1.5 mm, respectively. Absolute 3D reconstructions on the tissue show quantitatively good accuracy and the ability to spatially distinguish small tissue features (adipose strands of approximately 2.5 mm in width). The reconstruction algorithm developed provides strong evidence for the promise of surgical margin detection using microendoscopic EIT.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.