The iCare tonometer is a mobile alternative to GAT in a low to moderate IOP range, but our findings show a greater deviation than previously reported. In high IOP values, measurements with the iCare tonometer do not correlate well with GAT.
ObjectiveTo test agreement of two methods to measure intraocular pressure (IOP): rebound tonometry (RBT) and gold standard Goldmann applanation tonometry (GAT) in children with glaucoma.DesignObservational prospective cohort study.SettingTertiary paediatric glaucoma clinic at a single centre.Participants102 individuals attending a paediatric glaucoma clinic, mean (SD) age 11.85 (3.17), of whom 53 were male.Primary and secondary outcome measuresIntraocular pressure, central corneal thickness, child preference for measurement method.ResultsLimits of agreement for intraobserver and interobserver were, respectively, (−2.71, 2.98) mm Hg and (−5.75, 5.97) mm Hg. RBT frequently gave higher readings than GAT and the magnitude of disagreement depend on the level of IOP being assessed. Differences of 10 mm Hg were not uncommon. RBT was the preferred method for 70% of children.ConclusionsThere is poor agreement between RBT and GAT in children with glaucoma. RBT frequently and significantly overestimates IOP. However, ‘normal’ RBT readings are likely to be accurate and may spare children an examination under anaesthesia (EUA). High RBT readings should prompt the practitioner to use another standard method of IOP measurement if possible, or consider the RBT measurement in the context of clinical findings before referring the child to a specialist clinic or considering EUA.
Trabeculectomy following a failed XEN-45 implantation seems technically feasible; however, higher than expected number of cases with low IOP and avascular blebs were observed. This might be related to the cumulative effect of MMC in these cases.
Ophthalmologists show good levels of agreement with each other when using GAT, while technicians/nursing staff show better agreement when using the DCT. The DCT may be a better tonometer to use if permanently delegating IOP measurements to non-ophthalmologists, but measurements cannot be interchanged with the GAT.
There is an acceptable agreement between IOP measurements made with the reusable Goldmann tonometer prism and the disposable Tonosafe prism. Interobserver variability in IOP measurements within an outpatient setting is larger than that found within a research setting, and may be of a level that impacts on clinical decision-making.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.