Development of breast tumors is often accompanied by angiogenesis--the formation of new blood vessels. It is possible to image the effects of this process by tracking the uptake and washout of contrast agents in the vicinity of a lesion. In this article, a method for carrying out contrast subtraction mammography on a full-field digital mammography unit is described. Spectral measurements and modeling were performed to optimize the choice of x-ray target, kilovoltage and x-ray beam filtration for contrast digital mammography (CDM) on an available digital mammography system. Phantom studies were carried out to determine the sensitivity of CDM to iodine. Detection of iodine area densities of 0.3 mg/cm2 is possible for a circular object with a radius of 1.3 mm, which allows detection of uptake levels in the breast typically seen with cancer and some benign breast conditions. It was found that with a molybdenum anode x-ray tube, copper filtration could be used to effectively shape the x-ray spectrum to maximize the proportion of x rays with energies above the k edge of iodine. Simple logarithmic subtraction was found to be adequate in suppressing background signals dependent on the x-ray beam intensity and background thickness of the breast. The total x-ray dose from the procedure ranges between 1 and 3 mGy, similar to that from a conventional single view film mammogram. A clinical pilot study is currently being carried out to evaluate this technique.
Optimization of exposure parameters (target, filter, and kVp) in digital mammography necessitates maximization of the image signal-to-noise ratio (SNR), while simultaneously minimizing patient dose. The goal of this study is to compare, for each of the major commercially available full field digital mammography (FFDM) systems, the impact of the selection of technique factors on image SNR and radiation dose for a range of breast thickness and tissue types. This phantom study is an update of a previous investigation and includes measurements on recent versions of two of the FFDM systems discussed in that article, as well as on three FFDM systems not available at that time. The five commercial FFDM systems tested, the Senographe 2000D from GE Healthcare, the Mammomat Novation DR from Siemens, the Selenia from Hologic, the Fischer Senoscan, and Fuji's 5000MA used with a Lorad M-IV mammography unit, are located at five different university test sites. Performance was assessed using all available x-ray target and filter combinations and nine different phantom types (three compressed thicknesses and three tissue composition types). Each phantom type was also imaged using the automatic exposure control (AEC) of each system to identify the exposure parameters used under automated image acquisition. The figure of merit (FOM) used to compare technique factors is the ratio of the square of the image SNR to the mean glandular dose. The results show that, for a given target/filter combination, in general FOM is a slowly changing function of kVp, with stronger dependence on the choice of target/filter combination. In all cases the FOM was a decreasing function of kVp at the top of the available range of kVp settings, indicating that higher tube voltages would produce no further performance improvement. For a given phantom type, the exposure parameter set resulting in the highest FOM value was system specific, depending on both the set of available target/filter combinations, and on the receptor type. In most cases, the AECs of the FFDM systems successfully identified exposure parameters resulting in FOM values near the maximum ones, however, there were several examples where AEC performance could be improved.
Oblique incidence of x rays on an imaging detector causes blurring that reduces spatial resolution. For simple projection imaging this effect is small and often ignored. However, for breast tomosynthesis, the incidence angle can be larger (>20 degrees), leading to increased blur for some of the projections. The modulation transfer function (MTF) is measured for a typical phosphor-coupled flat-panel detector versus angular incidence of the x-ray beam for two x-ray spectra: 26 kV Mo/Mo and 40 kV Rh/Al. At an incidence angle of 40 degrees the MTF at 5 mm(-1) falls by 35% and 40% for each spectrum, respectively (and 65%/80% at 8 mm(-1)). Increasing the detector absorber thickness to achieve improved quantum efficiency will cause the blurring effect due to beam obliquity to become greater. The impact of this blur is likely to cause misregistration and increased relative noise in tomosynthesis reconstructed images.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.