This paper reports the first analysis of small molecules on the Agilent bio-analyser. The Bioanalyzer is a commercial lab-on-a-chip instrument designed for the analysis of DNA and proteins. We demonstrate that the instrument is suitable for analyses beyond its design specifications. Amphetamine, methamphetamine and pseudoephedrine were separated with a 50 mM borate and 50 mM sodium dodecyl sulfate (SDS) buffer at pH 9.66. The analytes were derivatised with fluorescein isothiocyanate (FITC) in 3 minutes with a heating block set at 90 C, reducing the typical time of 12 hours required for amine-labelling. Analytes were detected by LED-induced fluorescence (l ex ¼ 525 nm and l em ¼ 470 nm). Furthermore, five amphetamine analogues were baseline separated within 1 minute. An average limit of detection of 0.6 mg mL À1 and limit of quantification of 2.2 mg mL
À1were obtained for all analytes. These rapid analyses in conjunction with a fast and reliable derivatisation method with FITC demonstrate its potential use for the in-field analysis of samples of forensic significance.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.