A portable system using a direct contact for the measurement of ultrasonic velocity and broadband attenuation in bone is described (contact ultrasonic bone analyser, CUBA). Soft-tissue compensation is performed using an ultrasonic pulse-echo technique. CUBA has been successfully validated using reference materials, the precision of velocity and broadband attenuation measurements being typically 0.2% and 0.5% respectively. The clinical reproducibility has been assessed on the equine third metacarpal bone. The reproducibility of velocity measurement is typically 0.5% for cortical bone and 1% for cancellous bone. For broadband attenuation the reproducibility is typically 7% for cortical bone and 6% for cancellous bone. The lower reproducibility of the attenuation data is attributed to the high sensitivity to variations in the material properties of bone with small changes in transducer positioning. Coupling difficulties through an intact equine coat have been overcome and the system may now be assessed in the clinical environment, in both human and animal populations.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.