The aim of this study was to evaluate the impact of different file sizes on the accuracy of two electronic apex locators (EALs). Thirty extracted human single-rooted permanent mandibular incisors were used. A #10 K-file was inserted in the root canal until its end could be observed (using a light microscope) through the apical foramen. One millimetre was subtracted to establish working length (WL). Electronic readings were performed using MiniApex Locator or Root ZX II, from #10 K-file to #130 K-file. Statistical analysis was performed by two-way anova and Tukey test (P ≤ 0.05). From #60 to #130 K-file, observed differences were noted between the values obtained with both EALs and WL (P ≤ 0.05). The MiniApex Locator showed increased means when measurements were made with #50 to #70 and with #120 (P = 0.008) and #130 (P = 0.005) K-files. File sizes influenced the accuracy of EALs - the greater the instrumentation size, the higher mean differences compared to WL.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.