Result: Weight of the patients was <1 kg in 10%, and >2 kg in 75%. Catheters tip placement was as aimed, mostly superior cava vein (SCV), in 65.5%, and subclavian vein in remaining 34.5% and average duration of catheterization was 13.5 ± 9.6 and 8.9 ± 5.8 days, respectively. Blood sampling through both lumens was possible especially when the tip was at SCV. Reasons for catheter withdrawal were end of indication (45.9%), phlebitis/edema (21.3%), suspected infection (3.2%), accidental withdrawal (3.2%) and rupture of proximal end (3.2%). In three (4.9%) patients, withdrawal was due to serious complications (two cases of pleural leakage of infusion fluid and one breakage of the metallic guide). About 16.3% of the patients died with the catheter still in situ. Infection findings were positive tip culture (14.7%) and catheter-related sepsis (3.2%).
Conclusion:Insertion of PIDLCC is possible in neonates. The incidence of complications, mostly mechanical, requires careful evaluation of indications, and strict adherence to the procedure of insertion and manipulation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.