In this study, a frequency tuner system is developed for generating variable frequency magnetic fields for magnetic hyperthermia applications. The tuning device contains three specially designed phase lock loop devices that drive a resonant inverter working in the frequency band of 180–525 kHz. This tuner system can be adapted for other resonant inverters employed in the studies of ferrofluids with superparamagnetic nanoparticles. The performance of the whole system is also examined. Our findings were in agreement with the theoretical expectations of phase locking and frequency tuning. The system is tested for samples of a solid magnetic material of cylindrical shape and ferrofluids with differing concentrations of powdered magnetite. The observations indicate significant frequency changes of the magnetic field due to heating of the samples. These frequency variations can be a source of errors, which should not be neglected in experiments determining the specific absorption rate or power dissipated density.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.