In this paper, we present a lab-in-a-droplet bioassay strategy for a centrifugal microfluidics or lab-on-a-disc (LOAD) platform with three important advancements including density difference pumping, power to disc and bidirectional flow control. First, with the water based bioassay droplets trapped in a micro-channel filled with mineral oil, centrifugal force due to the density difference between the water and oil phases actuates droplet movement while the oil based medium remains stationary. Second, electricity is coupled to the rotating disc through a split-core transformer, thus enabling on-chip real-time heating in selected areas as desired and wireless programmable functionality. Third, an inertial mechanical structure is proposed to achieve bidirectional flow control within the spinning disc. The droplets can move back and forth between two heaters upon changing the rotational speed. Our platform is an essential and versatile solution for bioassays such as those involving DNA amplification, where localized temperature cycling is required. Finally, without the loss of generality, we demonstrate the functionality of our platform by performing real-time polymerase chain reaction (RT-PCR) in a linear microchannel made with PTFE (Teflon) micro-tubing.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.