Distal humerus fractures commonly occur in adults with low bone mineral density causing major technical challenges for orthopedic surgeons. Persian fixation method was introduced as a novel technique to stabilize small fragments in comminuted distal humerus fractures using a set of K-wires and a reconstruction plate. The present study aims to measure this technique's stiffness and stability of this technique and analyze the effect of influential parameters with numerical simulation and biomechanical testing on a cadaveric specimen. Validation of the finite element (FE) model was conducted based on results of experiments. The results indicated that Delta configuration mainly led to a higher stiffness in the case of axial loading and anterior bending compared to L configuration. Analyzing the influential factors of this technique suggests that changes in diameter and number of K-wires have a similarly significant effect on the construct stiffness while the height of plate had a slight influence. Also, the diameter of wires was the most effective parameter for implant failure, particularly in the 3-pin construct, which caused a reduction in failure risk by about 60%. The results revealed that the Persian fixation method would achieve suitable stability compared to the dual-plating technique.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.