Many patients with end-stage cardiomyopathy are now being implanted with Total Artificial Hearts (TAHs). We have observed individual cases of post-operative mechanical ventilator autocycling with a flow trigger, and subsequent loss of autocycling after switching to a pressure trigger. These observations prompted us to do a retrospective review of all TAH devices placed at our institution between August 2007 and May 2009. We found that in the immediate post-operative period following TAH placement, autocycling was present in 50% (5/10) of cases. There was immediate cessation of autocycling in all patients after being changed from a flow trigger of 2 L/minute to a pressure trigger of 2 cm H2O. The autocycling group was found to have significantly higher CVP values than the non-autocycling group (P = 0.012). Our data suggest that mechanical ventilator autocycling may be resolved or prevented by the use of a pressure trigger rather than a flow trigger setting in patients with TAHs who require mechanical ventilation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.