This paper describes two simple, specific, accurate, and precise methods for estimation of olopatadine hydrochloride (OLO) in tablet dosage form. The first method is a stability-indicating isocratic RP-HPLC method. The analysis is performed on an RP-18 column using 0.1% orthophosphoric acid (adjusted to pH 4.5 with triethylamine)-acetonitrile (75 + 25, v/v) mobile phase at a flow rate of 1 mL/min. Paracetamol (PAR) was selected as the internal standard. Retention times of OLO and PAR were 11.30 +/- 0.02 and 4.70 +/- 0.03 min, respectively. For the HPTLC method, precoated silica gel 60 F254 aluminum sheets were used as the stationary phase; the mobile phase was methanol-chloroform-ammonia (8 + 2 + 0.1, v/v/v). The detection of the analyte band was carried out at 301 nm, and its Rf value was 0.46 +/- 0.03. The analytical methods were validated according to International Conference on Harmonization guidelines. Linear regression analysis data for the calibration plots showed a good linear relationship between response and concentration in the range of 0.1-1 microg/mL and 0.1-0.9 microg/band for HPLC and HPTLC, respectively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.