The aim of this study was to evaluate the adequacy of single-frequency (sf-BIA) and multi-frequency bioelectrical impedance analyses (mf-BIA), in comparison with dual-energy x-ray absorptiometry (DXA), to evaluate body composition in maintenance haemodialysis (MHD) patients. Body composition of 27 adult MHD patients (9 f, 18 m), BMI 17.5-34.4 kg m(-2), was examined with DXA and BIA, with two different sf-BIA and 1 mf-BIA analysers. Biochemical markers of nutritional status and adequacy of dialytic treatment were also determined. Fat mass (FM) estimated by the different BIA analysers was found to be slightly but significantly higher than FM measured by DXA. In contrast, fat-free mass (FFM) obtained with BIA was found to be slightly but significantly lower than FFM DXA. No significant differences were found between LBM-DXA (that is FFM-DXA minus bone mass) and the different FFM BIA. The lowest mean prediction error versus DXA values was found with sf1BIA. In any case, a close correlation was found between all BIA values and DXA values, particularly for FFM. Furthermore, FFM and LBM results were significantly correlated with serum creatinine, which in MHD patients is an indicator of muscle mass. These results indicate that BIA can be used to evaluate body composition in MHD patients.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.