In the present work, numerical simulations were conducted for a typical end-to-side distal graft anastomosis to assess the effects of inducing secondary flow, which is believed to remove unfavourable flow environment. Simulations were carried out for four models, generated based on two main features of 'out-of-plane helicity' and 'spiral ridge' in the grafts as well as their combination. Following a qualitative comparison against in vitro data, various mean flow and hemodynamic parameters were compared and the results showed that helicity is significantly more effective in inducing swirling flow in comparison to a spiral ridge, while their combination could be even more effective. In addition, the induced swirling flow was generally found to be increasing the wall shear stress and reducing the flow stagnation and particle residence time within the anastomotic region and the host artery, which may be beneficial to the graft longevity and patency rates. Finally, a parametric study on the spiral ridge geometrical features was conducted, which showed that the ridge height and the number of spiral ridges have significant effects on inducing swirling flow, and revealed the potential of improving the efficiency of such designs.
Graft failure is currently a major concern for medical practitioners in treating Peripheral Vascular Disease (PVD) and Coronary Artery Disease (CAD). It is now widely accepted that unfavourable haemodynamic conditions play an essential role in the formation and development of intimal hyperplasia, which is the main cause of graft failure. This paper uses Computational Fluid Dynamics (CFD) to conduct a parametric study to enhance the design and performance of a novel prosthetic graft, which utilises internal ridge(s) to induce spiral flow. This design is primarily based on the identification of the blood flow as spiral in the whole arterial system and is believed to improve the graft longevity and patency rates at distal graft anastomoses. Four different design parameters were assessed in this work and the trailing edge orientation of the ridge was identified as the most important parameter to induce physiological swirling flow, while the height of the ridge also significantly contributed to the enhanced performance of this type of graft. Building on these conclusions, an enhanced configuration of spiral graft is proposed and compared against conventional and spiral grafts to reaffirm its potential benefits.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.