Venous catheters are increasingly used for chronic haemodialysis, with dual lumen catheters being the most commonly used as blood recirculation (REC%) is relatively low. The aim of this study was therefore to evaluate blood recirculation in dual lumen catheters, both well-functioning and malfunctioning, with reversed lumens. In our study, blood recirculation in well-functioning catheters with standard lumens is similar to that found in previous studies. However, when lumens are reversed, blood recirculation increases significantly (6.7 +/- 4 vs 19 +/- 11%, p < 0.001). REC% in malfunctioning catheters (10.8 +/- 2%) was higher than normal function (p < 0.05) but lower than reversed flow in normal catheters (p < 0.01). Therefore, inadvertent reversal of lumens in a well-functioning catheter increases REC% in a significant manner, thus worsening haemodialysis efficiency. We conclude that, in inflow failure catheters, lumens can be reversed because REC% is acceptable. However, inadvertent reversal of lumens in a well-functioning catheter increases REC% to a level which may compromise the adequacy of haemodialysis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.