(22R,S)Budesonide was isolated from human plasma by solid-phase extraction. Switching from reversed-phase conditions during sample application and washing to normal-phase conditions during elution afforded a very clean extract. Budesonide was derivatized with acetic anhydride to form the 21-acetyl derivative before analysis by reversed-phase liquid chromatography combined with thermospray mass spectrometry. Deuterium-labelled budesonide was used as internal standard. Standard samples prepared in human albumin solution were used for the calibration curve. An automated liquid chromatography/mass spectrometry system, allowing unattended overnight operation, was used for routine analysis. The recovery of budesonide from plasma was 88.9 +/- 5.9% (mean +/- SD) and the method was linear over the range 0.30-30 pmol (amount analysed), corresponding to plasma concentrations of 0.10-10 nmol l-1. Budesonide could be measured down to 0.10 nmol l-1 with a within-day variation of 10-18% (CV). The error was less than +/- 15% at 0.10 nmol l-1 and less than +/- 7% at concentrations of 0.20 nmol l-1 or higher. The total imprecision between days was 9% (CV) at a concentration of 0.30 nmol l-1.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.